Diluent blank
Measure background before adding particles.
A correct size standard can still produce a bad result when dilution, background contamination, bubbles, adsorption or sampling are uncontrolled.
| Method | ISO reference | Typical ISO abstract range | Important response factor |
|---|---|---|---|
| Light scattering liquid-borne | ISO 21501-2:2019 | 0.1–10 µm | Particle and liquid refractive indices influence optical-equivalent size |
| Light extinction liquid-borne | ISO 21501-3:2019 | 1–100 µm | Particle blocks or attenuates light; pharmaceutical liquid applications are included |
Measure background before adding particles.
Use calibrated devices within their useful range.
Define a repeatable resuspension method.
Avoid counting gas bubbles as particles.
Evaluate loss to containers and tubing.
Control settling and concentration drift.
A concentrated size standard can support a calibration preparation when diluted under a validated method. A ready-to-use count control can support routine performance monitoring. Neither automatically replaces the other.
Focus on certified mean, uncertainty and distribution.
Focus on defined number concentration, sampling volume and repeatable routine use.
Send the instrument, medium, target size, concentration, certificate and packaging requirements to Applied Physics for a technical recommendation.